title: "Product configurations and software options (Vicon Vantage system)"
canonical: "https://help.vicon.com/space/regs/12885915/Product%20configurations%20and%20software%20options%20(Vicon%20Vantage%20system)"
format: markdown
---
This topic provides information relating to the [EU Declaration of Conformity (Vicon Vantage system)](https://vicon-help.atlassian.net/wiki/spaces/regs/pages/12878946).

**Conformity of the Metrological Performance of CLASS 1 Products Manufactured in Accordance with **  
**Annex VII, Section 5 of the Medical Devices Directive 93/42/EEC of 14th June 1993. **  
**As amended by EU Council Directive 2007/47/EC of 5th September 2007.**

We, Vicon Motion Systems Limited  
Unit 6, Oxford Industrial Park  
Mead Road, Yarnton, Oxford, OX5 1QU  
UNITED KINGDOM

declare that the VICON Vantage System manufactured by VICON MOTION SYSTEMS LIMITED has been tested prior to shipment and meets the following metrological performance:

**Measurement criteria**

- Supporting software Nexus 2.10 or later, Tracker 3.8 or later, Shogun 1.4 or later.
- Using a minimum of four cameras, resolution of the distance between the centers of two static 14mm spherical markers located within a volume no less than 4mx4mx1.5m to within 1mm Mean; 1mm Standard Deviation; sample size no less than 1,000.

**Analogue Digital Conversion**

Resolution to ± 10 mV mean and ± 10 mV (1 Standard Deviation).

** ****Synchronization**

Difference within one video frame.