Ensure the FLIR cameras are exactly synchronized to the Vicon system. To verify this:

1. Set the Vicon system and FLIR camera to the same frame rate eg. 100 Hz.
2. Position the Active Wand where it can be seen in both the video and by a Vicon optical camera.
3. Set the Active Wand to Strobe mode.
4. Start capturing.
5. Turn the wand on.
6. Turn the wand off.
7. Stop capturing.
8. Transfer the file and review it by stepping through the frames.
9. Find the video frame where the wand is switched on and check that is the same frame where the wand appears in the optical camera.

> ⚠️ **Synchronization issues**  
> ⚠️ If you encounter synchronization issues, try re-synchronizing the FLIR video camera.   
> ⚠️ If the issue persists, check for errors in the log and if these don't enable you to resolve the issue, please contact [Vicon Support](mailto:support@vicon.com).